Lung Cancer Screening CT Phantom LSCT001 (PH-8)

SKU: N/A

Chest phantom for standardization studies in low dose lung cancer CT screening / Anthropomorphic structure provides life-like images allowing operators visual evaluation

Add To Quote
SKU: N/A Category: Brand:

Description

Realizing Education That Inspires Through Manufacturing

Founded in 1891, Kyoto is the world’s only manufacturer of anatomical models, simulators, and imaging phantoms. Made in Kyoto, Japan and trusted in over 60 countries around the world. Passing on cultural heritage through technologies for restoring and replicating cultural assets.

Features:

  • Simulated GGO type tumors with different sizes and HU numbers are prepared in the vicinity of three main sections of bilateral lungs
  • Dosimeter holder on the central axis of the phantom allows housing a pencil type ion chamber. 8 step cylindrical linearity phantom to control density curve as a scale can be attached to the chest phantom base.

Applications:

  • CT image quality evaluation
  • Dosimetry
  • Evaluation of density curve

Anatomy

  • Bones
  • Lungs
  • Mediastinum

Simulated tumors at three lung areas:

  • Apical portion of the lungs
  • Bifurcation of the trachea
  • Base of lungs

Set Includes

1 chest phantom 1 adjustment base
1 8 step linearity phantom 1 set of sample images
1 urethane cylinder manual

Materials

Chest wall: human tissue substitute
Bones: synthetic bones
Alveoli: styrene foam and urethane foam

Dimensions:

Phantom size:
W44×H69.4 cm
W17.3×H27.3 in

Simulated tumors HU contrast with the
lung back ground
Size Materials
tumors in the right lung HU=100 4, 6, 8, 10, 12 mm dia.
0.16, 0.24, 0.32, 0.39, 0.47 in dia.
urethane resin
tumors in the left lung HU=270 2, 4, 6, 8, 10 mm dia.
0.08, 0.16, 0.24, 0.32, 0.39 in dia.
urethane resin

Linearity phantom targets

HU contrast with the
lung back ground
Materials
A -1000 Air
B -850 polyurethane
C -600 polyurethane
D -400 polyurethane
E -200 polyurethane
F 100 polyurethane
G 250 Bakelite
H 350 polyacetal resin